Textile Far Infrared Radiation Temperature Rise Tester



Instrument Application

Used for various textile products including fibers, yarns, fabrics, nonwovens and their derivatives to determine far infrared properties through temperature rise tests.

Compliance Standards

GB/T30127 4.2 ,T/CAS115.5-2021

Test Principle

After irradiating specimens with a constant-intensity far infrared radiation source for specified duration, measure the temperature elevation on the specimen's test surface.

 Product Features

Thermal insulation baffle: Front-mounted heat shield improves test accuracy and reproducibility

Fully automatic measurement: Tests initiate automatically when lid is closed

Equipped with Matsushita (Panasonic) power meter for real-time power monitoring

Uses Omega sensors and transmitters

Three specialized specimen holders (yarn/fiber/fabric)

Optical modulation technology eliminates environmental radiation interference

Technical Parameters

Parameter Item

Parameter Value

Specimen Holder

500mm distance from radiation source to specimen surface

Radiation Source

Main wavelength 5μm~14μm, radiation power 150W

Specimen Irradiation Area

φ60~φ80mm

Temperature Range & Accuracy

15℃~50℃, ±0.1℃ precision, ≤1s response time

Specimen Types

  • Yarns: ≥60mm square metal frame
  • Fibers: φ60mm×30mm open cylindrical container
  • Fabrics: φ≥60mm