G506 Mask Automatic Filter Performance Tester


The latest G506 automatic filter performance tester uses the most advanced technology and design concepts, to meet a variety of test standards and the latest National standard GB / T 32610.Only relatively little operation is required to complete all the function and operation. Hardware performance of G506 mask automatic filter performance tester is stable and excellent in repeatability. Software design meet with people's thinking mode, the interface is user-friendly, and the functional and parameter settings are very reasonable. It is convenient to user for academic research and product development.

Application

G506 Automatic Filter Performance Tester is used for testing professional filter material and filtration efficiencyof a wide range of air filters including HEPA and ULPA filter, respirator, filter cartridge and protective clothing test. It also can use salt or oil mist to test filter media. Rapid acquisition of filter media performance evaluation parameters, and accurate evaluation of filter material quality, performance, life and so on. It is convenient to user for academic research and product development of new filter materials, product quality control, material performance verification.

Standards

ASTM F2100: Standard Specification for Performance of Materials Used in Medical Face Masks

EN149: Respiratory protective devices - Filtering half masks to protect against particles - Requirements, testing, marking

NIOSH-42 CFR84: NIOSH Guide to the Selection and Use of Particulate Respirators(或 42 CFR Part 84 - Approval of Respiratory Protective Devices)

Feature

Complies to more standards than any other filter tester

Salt and oil testing in one unit

Highly consistent test results

New user-interface and data recording options

Customizable fully automated testing sequences

Ergonomic design

Reduced cost of ownership with new serviceable photometers

Worldwide service and support

Parameters

ItemsParameter
ApllicationLoad and quality control
Max Efficiency99.999%
Aerosol TypeDOP/PAO/DEHS/Paraffi/Others Oil Aerosol、Salt
Aerosol GenerationAtomizing
CMD0.2μm(Oil)or 0.75um(NaCl)
GSD<1.6(Oil)or <1.86(NaCl)
Flow Rate15-100L/min
Resistance0-150mm H2O(0-1470Pa)
Particle DetectScatter Photometer
Measurement Time10 Sec.
Data ReportLCD,RS-232
OperationIndependent tewster、automatic production line
Standard compliantUS 42 CFR part 84.EN 143.JMOL,ISO 23328-1 ISO 16900-3
Power Supply115VAC、60Hz,10A或230VAC,50Hz,5A
Compressor Air80psig,7scfm(550kPa,198std.L/min)
Dimension70cm x 70cm x 147cm(28in, x 28in.x 58in.)